Features of Controlling Tower

BR100-C1-5L.png

Multiple Control Function

  • Associated control: temperature, stirring speed, pH, dissolved oxygen, defoaming, airflow, tank pressure, feed amount and other direct parameters.

  • Remote monitoring: Suitable for smartphones, tablets, laptops and PC for instant remote monitoring.

  • Step control: Users can switch directly between manual control and step control and set at least 10-step control program for temperature, agitation, PH, DO and feeding.

Smart PID control

  • Temperature, stirring speed, Ph, DO and other parameters in the tank can be controlled by subsection curve setting according to the process requirements.

  • Temperature Control: Adopt PID automatic control mode; SS316L jacket base for electric heating and cold water for cooling.

  • Dissolved oxygen control: equipped with DO electrode and shielded conductor, DO is associated with stirring motor speed, feeding, airflow, tank pressure and other parameters.

  • PH value control: Intelligent PID system control built-in control peristaltic pumps for acid-alkali adjustment, with high control precision.

  • Antifoam control: Foams are automatically detected and the peristaltic pump is controlled to add defoamer in proportion to time.

  • Feeding control: Peristaltic pump switch control (manual, automatic, and off); accumulation volume control; coupled with other parameters (DO, pH, etc).

截屏--发酵计划.jpg
截屏--Temp.jpg

Standard Glass Vessel

pexels-ivan-samkov-9629704.jpg
  • Equipped with an easy-to detach condenser around the air outlet, for easy cleaning, no effusion and no contaminating bacteria.

  • Standard sensors port for precise measurement and control of temperature, pH, DO, Antifoam, feed, aeration and pressure.

  • Magnetic drive stirring for good sealing performance.

  • Compact design for saving critical laboratory space..

  • Standard with flat or pitched-blade; other types of impellers are available

  • Configured one sampling port.

  • Equipped with a filter to prevent contaminated air from getting to the system.

  • Tempered high borosilicate glass 3.3 for vessel and SS316L for the heating plate.

  • Perfectly suited for research and development, available in 1L, 2L, 3L, 5L, 7L, 10L and 15L.

IMG_0757.jpg
IMG_0756.jpg
Glass-Vessel.png

Upgrading Option List

Standard Type Specifications

Model
BR100-C1
BR100-C2
Controller
One console for one vessel
One console for two vessels
Display / Operation
Touch panel 10" / Touch screen
Touch panel 10" / Touch screen
Gas inlet
1 overlay with 4 deep gas inlets
1 overlay with 4 deep gas inlets for each vessel
USB
1 x RS485
2 x RS485
Vessel
1L / 2L / 3L / 5L / 7L / 10L
2 x 1L / 2L / 3L / 5L / 7L / 10L
Working volume
0.8L / 2L / 2.6L / 4.4L / 6L / 8.4L
2 x 0.8L / 2L / 2.6L / 4.4L / 6L / 8.4L
Stirring speed
5 - 1000 rpm
5 - 1000 rpm
Motor power (Kw)
0.18/0.4/0.4/0.4/0.4/0.75
0.18/0.4/0.4/0.4/0.4/0.75
Pressure
0.10Mpa (Max 0.15Mpa)
0.10Mpa (Max 0.15Mpa)
Material
SS316L & 3.3 high borosilicate glass
SS316L & 3.3 high borosilicate glass
Serilization
Autoclave
Autoclave
Feeding pumps
4 peristaltic pumps
8 peristaltic pumps