LAB1ST PSC-G Series Pharmaceutical Stability Test Chamber with illumination provides complete stability testing solutions for pharmaceuticals requiring light exposure studies.
With precise temperature control from 0 °C to 65 °C (illumination range: 15–50°C) and humidity control from 25% to 95% RH ±3% RH, this series supports accelerated, long-term, and photostability testing.
The integrated illumination system with stepless dimming and built-in light intensity monitoring makes it ideal for modern pharmaceutical laboratories.
Features
-
Energy-Efficient & Eco-Friendly Design
-
Fluorine-free refrigeration system for high efficiency, low energy consumption, and environmental sustainability.
-
Advanced Control & Monitoring System
-
Standard Illuminance Monitor with top-mounted lighting and manual stepless dimming
-
5-inch Touchscreen Programmable Controller (7-inch for PSC-1500L-G) supporting 100 programs, 1000 segments, and 999 cycles, with maximum duration of 99 hours 59 minutes per segment
-
Dual International Brand Compressors with automatic switching for continuous, uninterrupted operation (Note: PSC-80L model equipped with one independent fully-enclosed compressor)
-
Advanced PID Auto-Tuning ensures precise and stable temperature and humidity control
-
Microcomputer Servo Controller with digital temperature and humidity display for accurate real-time monitoring
-
Frost-Free Continuous Operation technology eliminates defrost cycles, preventing temperature and humidity fluctuations during critical tests
-
Bilingual Interface (Chinese/English) with real-time curve display
-
Advanced Control & Monitoring System
-
High-Grade 304 Stainless Steel Interior featuring smooth, radiused corners for effortless cleaning and optimal contamination control
-
Advanced Airflow Duct System delivering exceptional air circulation uniformity throughout the entire workspace
-
25mm Access Port conveniently located on the left side for easy sensor integration and monitoring
-
Reliable Components & Sensors
-
Internationally Renowned Brand Components for temperature/humidity controllers, compressors, and circulating fans.
-
High-Temperature Resistant Humidity Sensor eliminates frequent replacement of wet wicks.
-
Data Recording & Communication
-
Standard RS485 Communication & Embedded Printer for real-time data logging and remote monitoring.
-
PC Software Support for programming, monitoring, and remote control.
-
Fault Diagnosis Display for clear and immediate error identification.
-
Comprehensive Safety & Alarm System
-
Independent Temperature Limit Alarm with audible and visual alerts for immediate hazard notification
-
Multi-Level Protection System including compressor overheat, fan overheat, over-temperature, over-pressure, overload, and water shortage protection
-
Dual-Direction Alarms for both high/low temperature and humidity deviations from set parameters
-
Password-Protected Screen Lock to prevent unauthorized access and operation
-
Hierarchical Access Management featuring multi-user accounts with three permission levels (Administrator, Operator, Visitor), supporting personalized login with detailed operation logs for audit trail compliance
-
Automatic Safety Shutdown that stops air circulation and heating when chamber door is opened, preventing temperature overshoot and ensuring operator safety
-
Compliance with International Standards
-
Comprehensive Testing Protocol Support – Fully compliant with ICH and other international regulatory guidelines for pharmaceutical stability testing, including:
-
Accelerated Testing: 40 °C ±2 °C / 75% RH ±5% RH or 30 °C ±1 °C / 60% RH ±5% RH (180 days)
-
Long-Term Testing: 25 °C ±2 °C / 60% RH ±5% RH or 30 °C ±2 °C / 60% RH ±5% RH (365 days)
-
Accelerated Testing for Semi-Permeable Packaging (e.g., LDPE infusion bags, plastic ampoules, ophthalmic containers): 40 °C ±2 °C / 25% RH ±5% RH
-
Long-Term Testing for Semi-Permeable Packaging: 25 °C ±2 °C / 40% RH ±5% RH or 30 °C ±2 °C / 35% RH ±5% RH
-
Strong Light Exposure Test: 4500 ±500 Lx (10 days)
-

-
Specical Features for PSC-1500L-G
-
7-inch LCD Touchscreen Controller with graphical interface and curve display
-
Standard USB Interface & Wired Monitoring and Alarm System

Specifications
Model | PSC-150L-G | PSC-250L-G | PSC-500L-G | PSC-800L-G | PSC-1000L-G | PSC-1500L-G |
|---|---|---|---|---|---|---|
Rated Power | 2950 W | 3200 W | 3910 W | 3600 W | 3700 W | 6000 W |
Chamber Volume | 150 L | 250 L | 500 L | 800 L | 1000 L | 1500 L |
Inner Dimensions (W×D×H, mm) | 550×405×670 | 600×500×830 | 670×725×1020 | 730×750×1480 | 730×890×1480 | 1550×590×1650 |
External Dimensions (W×D×H, mm) | 690×805×1530 | 740×890×1680 | 850×1100×1930 | 870×1085×2065 | 870×1220×2065 | 2110×890×2000 |
Standard Shelves/Trays | 3 pcs | 3 pcs | 4 pcs | 4 pcs | 4 pcs | 4 pcs |
Operating Power (approx.) | ~680 W | ~750 W | ~1400 W | ~2880 W | ~2960 W | ~4800 W |
Power Supply | AC220V 50HZ | AC380V 50HZ | ||||
Temp. Fluctuation/Uniformity | ±0.5℃/±2℃ | ±0.5℃/±2℃~±3℃ |
Safety Devices | Compressor overheat protection, Fan overheat protection, Over-temperature protection, Compressor overpressure protection, Overload protection, Water shortage protection |
Communication Interface | 485 interface |
Test Port | 25mm test port |
Built-in Printer | Yes |
Illumination Intensity/Error | 0~6000LX Adjustable ≤±500LX (Stepless dimming) |
Timer Range | Each segment 1~99 hours |
Temp. & Humidity Control Method | Balanced temperature and humidity control method |
Refrigeration System/Method | Two independent sets of fully hermetic compressors, auto-switch, international brand |
Controller | Programmable touch screen controller |
Sensor | Pt100 platinum resistor; Capacitive humidity sensor |
Ambient Operating Temp. | +5~30℃ |
Humidity Range/Deviation | 25~95%RH, ±3%RH |
Temp. Control Range | 0~65℃ (With illumination 15~50℃) |
Options
-
Paperless Recorder (Universal Type)
-
Wired Monitoring and Alarm System
-
USB Interface for Data Export
-
UV Sterilization System
-
International Brand SDR100 Paperless Recorder (2/4/6 Channels)
-
3Q Validation and Calibration Services
-
Shelf-Mounted Lighting (per shelf)
-
UV Control with 320–400 nm UV lamps and UV irradiance monitor (refer to Appendix 1)

.png)